SciELO - Scientific Electronic Library Online

 
vol.31 issue3Functional results of open wedge osteotomies in lower limb angular deformities in teenagersFemur fracture in a patient with dog-leg deformity secondary to arthrogryposis. Description of a unique osteosynthesis method author indexsubject indexsearch form
Home Pagealphabetic serial listing  

Services on Demand

Journal

Article

Indicators

Related links

  • Have no similar articlesSimilars in SciELO

Share


Acta ortopédica mexicana

Print version ISSN 2306-4102

Abstract

ALMAZAN-DIAZ, A et al. Intraarticular lesions associated to a tibial fixation device in anterior cruciate ligament reconstruction. Acta ortop. mex [online]. 2017, vol.31, n.3, pp.145-147. ISSN 2306-4102.

Anterior cruciate ligament reconstruction is a demanding procedure; complications may arise at any of its stages. This is the report of three cases in which the tibial fixation devices (IntraFix and Bio-IntraFix) led to intraarticular lesions. The complications were detected in one of the cases while the patient was still in the operating room and in the remaining two during the follow-up. All the patients had a stable knee, but all of them sustained chondral lesions resulting from the inadequate placement of the implants used for tibial fixation. Several suggestions to avoid the complications stemming from these devices are provided.

Keywords : Intrafix; divergence; complication; ACL reconstruction.

        · abstract in Spanish     · text in Spanish     · Spanish ( pdf )