<?xml version="1.0" encoding="ISO-8859-1"?><article xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance">
<front>
<journal-meta>
<journal-id>1405-9940</journal-id>
<journal-title><![CDATA[Archivos de cardiología de México]]></journal-title>
<abbrev-journal-title><![CDATA[Arch. Cardiol. Méx.]]></abbrev-journal-title>
<issn>1405-9940</issn>
<publisher>
<publisher-name><![CDATA[Instituto Nacional de Cardiología Ignacio Chávez]]></publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id>S1405-99402014000400003</article-id>
<article-id pub-id-type="doi">10.1016/j.acmx.2013.10.013</article-id>
<title-group>
<article-title xml:lang="es"><![CDATA[Cierre percutáneo de la comunicación interauricular con el dispositivo Amplatzer®: 15 años de experiencia]]></article-title>
<article-title xml:lang="en"><![CDATA[Percutaneous closure of atrial septal defects with the Amplatzer® device: 15 years of experience]]></article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author">
<name>
<surname><![CDATA[Zabal-Cerdeira]]></surname>
<given-names><![CDATA[Carlos]]></given-names>
</name>
<xref ref-type="aff" rid="A01"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname><![CDATA[García-Montes]]></surname>
<given-names><![CDATA[José Antonio]]></given-names>
</name>
<xref ref-type="aff" rid="A01"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname><![CDATA[Sandoval-Jones]]></surname>
<given-names><![CDATA[Juan Pablo]]></given-names>
</name>
<xref ref-type="aff" rid="A01"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname><![CDATA[Calderón-Colmenero]]></surname>
<given-names><![CDATA[Juan]]></given-names>
</name>
<xref ref-type="aff" rid="A02"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname><![CDATA[Patiño-Bahena]]></surname>
<given-names><![CDATA[Emilia]]></given-names>
</name>
<xref ref-type="aff" rid="A02"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname><![CDATA[Juanico-Enríquez]]></surname>
<given-names><![CDATA[Antonio]]></given-names>
</name>
<xref ref-type="aff" rid="A02"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname><![CDATA[Buendía-Hernández]]></surname>
<given-names><![CDATA[Alfonso]]></given-names>
</name>
<xref ref-type="aff" rid="A02"/>
</contrib>
</contrib-group>
<aff id="A01">
<institution><![CDATA[,Instituto Nacional de Cardiología Ignacio Chávez Departamento de Hemodinámica ]]></institution>
<addr-line><![CDATA[México Distrito Federal]]></addr-line>
<country>México</country>
</aff>
<aff id="A02">
<institution><![CDATA[,Instituto Nacional de Cardiología Ignacio Chávez Departamento de Cardiología Pediátrica ]]></institution>
<addr-line><![CDATA[México Distrito Federal]]></addr-line>
<country>México</country>
</aff>
<pub-date pub-type="pub">
<day>00</day>
<month>12</month>
<year>2014</year>
</pub-date>
<pub-date pub-type="epub">
<day>00</day>
<month>12</month>
<year>2014</year>
</pub-date>
<volume>84</volume>
<numero>4</numero>
<fpage>250</fpage>
<lpage>255</lpage>
<copyright-statement/>
<copyright-year/>
<self-uri xlink:href="http://www.scielo.org.mx/scielo.php?script=sci_arttext&amp;pid=S1405-99402014000400003&amp;lng=en&amp;nrm=iso"></self-uri><self-uri xlink:href="http://www.scielo.org.mx/scielo.php?script=sci_abstract&amp;pid=S1405-99402014000400003&amp;lng=en&amp;nrm=iso"></self-uri><self-uri xlink:href="http://www.scielo.org.mx/scielo.php?script=sci_pdf&amp;pid=S1405-99402014000400003&amp;lng=en&amp;nrm=iso"></self-uri><abstract abstract-type="short" xml:lang="es"><p><![CDATA[Introducción: El cierre percutáneo de la comunicación interauricular es una alternativa al tratamiento quirúrgico. Objetivos: Comunicar los resultados del cierre percutáneo de la comunicación interauricular con el dispositivo Amplatzer®. Método: Se incluyeron los pacientes sometidos a hemodinamia de septiembre de 1997 a diciembre de 2011. Resultados: Se procedió al cierre del defecto en 721 pacientes y en 85 no se cerró por considerarse no apto anatómicamente. El diámetro expandido del defecto fue de 23 ± 6.7 mm (límites, 5-42). Todos los dispositivos se colocaron con éxito. En 15 casos (2.1%) se cambió el dispositivo por considerarse inestable y en 6 (0.8%) se presentó embolización en las primeras 24 h, 4 fueron rescatados y recolocados (éxito final con intención de tratamiento 719/806-89.2%). El control inmediato mostró oclusión completa en 247 pacientes (34.3%), fuga intradispositivo en 395 (54.9%), ligera en 75 (10.4%) y moderada en 2 (0.3%). Se hace seguimiento en 626 pacientes (87.1%) durante 33 ± 27.5 meses, presentándose cefalea en 54 (8.6%), arritmias supraventriculares nuevas en 10 (1.6%), complicaciones en el sitio de punción en 4 (0.6%) y accidente vascular cerebral en uno (0.1%). Conclusión: El diseño del dispositivo Amplatzer® permite el cierre de la comunicación interauricular con una técnica sencilla y con seguridad para el paciente.]]></p></abstract>
<abstract abstract-type="short" xml:lang="en"><p><![CDATA[Introduction: Percutaneous closure of atrial septal defects is an alternative to surgical treatment. Objectives: We report the results of percutaneous closure of atrial septal defects with the Amplatzer® device. Method: We include patients taken to the catheterization laboratory from September 1997 to December 2011. Results: We proceeded with defect closure in 721 patients and in 85 the defect was considered not suitable for closure. The stretched diameter of the defect was 23 ± 6.7 mm (limits, 5-42). All devices were positioned successfully. In 15 cases (2.1%) the device was changed due to instability and in 6 (0.8%) the device embolized within the first 24 h, 4 were retrieved and repositioned (final success with intention to treat 719/806-89.2%). Immediate control showed complete closure in 247 patients (34.3%), leak through the device in 395 (54.9%), mild residual leak in 75 (10.4%) and moderate in 2 (0.3%). We have follow-up in 626 patients (87.1%) for 33 ± 27.5 months, with headache in 54 (8.6%), new supraventricular arrhythmia in 10 (1.6%), puncture site complications in 4 (0.6%), and stroke in one (0.1%). Conclusion: The design of the Amplatzer® device allows effective closure of atrial septal defects with a simple technique and patient safety.]]></p></abstract>
<kwd-group>
<kwd lng="es"><![CDATA[Comunicación interauricular]]></kwd>
<kwd lng="es"><![CDATA[Cierre percutáneo]]></kwd>
<kwd lng="es"><![CDATA[Dispositivo Amplatzer®]]></kwd>
<kwd lng="es"><![CDATA[Cardiopatías congénitas]]></kwd>
<kwd lng="es"><![CDATA[Cardiología intervencionista]]></kwd>
<kwd lng="es"><![CDATA[México]]></kwd>
<kwd lng="en"><![CDATA[Atrial septal defect]]></kwd>
<kwd lng="en"><![CDATA[Percutaneous closure]]></kwd>
<kwd lng="en"><![CDATA[Amplatzer® device]]></kwd>
<kwd lng="en"><![CDATA[Congenital heart disease]]></kwd>
<kwd lng="en"><![CDATA[Interventional cardiology]]></kwd>
<kwd lng="en"><![CDATA[Mexico]]></kwd>
</kwd-group>
</article-meta>
</front><body><![CDATA[  	    <p align="justify"><font face="verdana" size="4">Investigaci&oacute;n cl&iacute;nica</font></p>      <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="center"><font face="verdana" size="4"><b>Cierre percut&aacute;neo de la comunicaci&oacute;n interauricular con el dispositivo Amplatzer<sup>&#174;</sup>: 15 a&ntilde;os de experiencia</b></font></p>      <p align="center"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="center"><font face="verdana" size="3"><b>Percutaneous closure of atrial septal defects with the Amplatzer<sup>&#174;</sup> device: 15 years of experience</b></font></p>      <p align="center"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="center"><font face="verdana" size="2"><b>Carlos Zabal&#45;Cerdeira<sup>a,</sup>*, Jos&eacute; Antonio Garc&iacute;a&#45;Montes&ordf;, Juan Pablo Sandoval&#45;Jones&ordf;, Juan Calder&oacute;n&#45;Colmenero<sup>b</sup>, Emilia Pati&ntilde;o&#45;Bahena<sup>b</sup>, Antonio Juanico&#45;Enr&iacute;quez<sup>b</sup>, Alfonso Buend&iacute;a&#45;Hern&aacute;ndez<sup>b</sup></b></font></p>  	    <p align="center"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><sup><i>a</i></sup> <i>Departamento de Hemodin&aacute;mica, Instituto Nacional de Cardiolog&iacute;a Ignacio Ch&aacute;vez, M&eacute;xico, D.F., M&eacute;xico.</i></font></p>  	    ]]></body>
<body><![CDATA[<p align="justify"><font face="verdana" size="2"><i><sup>b</sup> Departamento de Cardiolog&iacute;a Pedi&aacute;trica, Instituto Nacional de Cardiolog&iacute;a Ignacio Ch&aacute;vez, M&eacute;xico, D.F., M&eacute;xico.</i></font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>* Autor para correspondencia:</b>    <br> 	Juan Badiano n.&#176; 1, Col. Secci&oacute;n XVI,    <br> 	Tlalpan, 14080 M&eacute;xico, D.F.    <br> 	Tel&eacute;fono.: &#43;52 (55) 5573 2911,    <br> 	ext. 1336, 1276, fax: &#43;52 (55) 5485 1568.    <br> 	Correos electr&oacute;nicos: <a href="mailto:zabal@cardiologia.org.mx">zabal@cardiologia.org.mx</a>, <a href="mailto:carlos@zabal.com.mx">carlos@zabal.com.mx</a> (C. Zabal&#45;Cerdeira).</font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2">Recibido el 8 de agosto de 2013    ]]></body>
<body><![CDATA[<br> 	Aceptado el 21 de octubre de 2013</font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>Resumen</b></font></p>  	    <p align="justify"><font face="verdana" size="2"><b><i>Introducci&oacute;n:</i></b> El cierre percut&aacute;neo de la comunicaci&oacute;n interauricular es una alternativa al tratamiento quir&uacute;rgico.</font></p>  	    <p align="justify"><font face="verdana" size="2"><b><i>Objetivos:</i></b> Comunicar los resultados del cierre percut&aacute;neo de la comunicaci&oacute;n interauricular con el dispositivo Amplatzer<sup>&#174;</sup>.</font></p>  	    <p align="justify"><font face="verdana" size="2"><b><i>M&eacute;todo:</i></b> Se incluyeron los pacientes sometidos a hemodinamia de septiembre de 1997 a diciembre de 2011.</font></p>  	    <p align="justify"><font face="verdana" size="2"><b><i>Resultados:</i></b> Se procedi&oacute; al cierre del defecto en 721 pacientes y en 85 no se cerr&oacute; por considerarse no apto anat&oacute;micamente. El di&aacute;metro expandido del defecto fue de 23 &#177; 6.7 mm (l&iacute;mites, 5&#45;42). Todos los dispositivos se colocaron con &eacute;xito. En 15 casos (2.1&#37;) se cambi&oacute; el dispositivo por considerarse inestable y en 6 (0.8&#37;) se present&oacute; embolizaci&oacute;n en las primeras 24 h, 4 fueron rescatados y recolocados (&eacute;xito final con intenci&oacute;n de tratamiento 719/806&#45;89.2&#37;). El control inmediato mostr&oacute; oclusi&oacute;n completa en 247 pacientes (34.3&#37;), fuga intradispositivo en 395 (54.9&#37;), ligera en 75 (10.4&#37;) y moderada en 2 (0.3&#37;). Se hace seguimiento en 626 pacientes (87.1&#37;) durante 33 &#177; 27.5 meses, present&aacute;ndose cefalea en 54 (8.6&#37;), arritmias supraventriculares nuevas en 10 (1.6&#37;), complicaciones en el sitio de punci&oacute;n en 4 (0.6&#37;) y accidente vascular cerebral en uno (0.1&#37;).</font></p>  	    <p align="justify"><font face="verdana" size="2"><b><em>Conclusi&oacute;n:</em></b> El dise&ntilde;o del dispositivo Amplatzer<sup>&#174;</sup> permite el cierre de la comunicaci&oacute;n interauricular con una t&eacute;cnica sencilla y con seguridad para el paciente.</font></p>      <p align="justify"><font face="verdana" size="2"><b>Palabras clave:</b> Comunicaci&oacute;n interauricular; Cierre percut&aacute;neo; Dispositivo Amplatzer<sup>&#174;</sup>; Cardiopat&iacute;as cong&eacute;nitas; Cardiolog&iacute;a intervencionista; M&eacute;xico.</font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    ]]></body>
<body><![CDATA[<p align="justify"><font face="verdana" size="2"><b>Abstract</b></font></p>  	    <p align="justify"><font face="verdana" size="2"><b><i>Introduction:</i></b> Percutaneous closure of atrial septal defects is an alternative to surgical treatment.</font></p>  	    <p align="justify"><font face="verdana" size="2"><b><i>Objectives:</i></b> We report the results of percutaneous closure of atrial septal defects with the Amplatzer<sup>&#174;</sup> device.</font></p>  	    <p align="justify"><font face="verdana" size="2"><b><i>Method:</i></b> We include patients taken to the catheterization laboratory from September 1997 to December 2011.</font></p>  	    <p align="justify"><font face="verdana" size="2"><b><i>Results:</i></b> We proceeded with defect closure in 721 patients and in 85 the defect was considered not suitable for closure. The stretched diameter of the defect was 23 &#177; 6.7 mm (limits, 5&#45;42). All devices were positioned successfully. In 15 cases (2.1&#37;) the device was changed due to instability and in 6 (0.8&#37;) the device embolized within the first 24 h, 4 were retrieved and repositioned (final success with intention to treat 719/806&#45;89.2&#37;). Immediate control showed complete closure in 247 patients (34.3&#37;), leak through the device in 395 (54.9&#37;), mild residual leak in 75 (10.4&#37;) and moderate in 2 (0.3&#37;). We have follow&#45;up in 626 patients (87.1&#37;) for 33 &#177; 27.5 months, with headache in 54 (8.6&#37;), new supraventricular arrhythmia in 10 (1.6&#37;), puncture site complications in 4 (0.6&#37;), and stroke in one (0.1&#37;).</font></p>  	    <p align="justify"><font face="verdana" size="2"><b><i>Conclusion:</i></b> The design of the Amplatzer<sup>&#174;</sup> device allows effective closure of atrial septal defects with a simple technique and patient safety.</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>Keywords:</b> Atrial septal defect; Percutaneous closure; Amplatzer<sup>&#174;</sup> device; Congenital heart disease; Interventional cardiology; Mexico.</font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>Introducci&oacute;n</b></font></p>  	    <p align="justify"><font face="verdana" size="2">El tratamiento del defecto septal auricular con cirug&iacute;a es un m&eacute;todo con eficacia ampliamente comprobada y con una morbimortalidad muy baja, aunque no est&aacute; exento de problemas tanto a corto como a largo plazo como defectos auriculares residuales, arritmias persistentes y enfermedad mioc&aacute;rdica<sup>1</sup> que dependen mucho de la edad de la cirug&iacute;a<sup>2</sup>. Desde que King y Mills<sup>3</sup> realizaron el primer intento de cierre de comunicaci&oacute;n interauricular transcateterismo, se han desarrollado varios tipos de dispositivos que en su mayor&iacute;a se basan en el principio de 2 discos opuestos para cerrar el defecto<sup>4&#150;13</sup>. El dispositivo Amplatzer<sup>&#174;</sup> se basa en el principio de distender el defecto septal ocluy&eacute;ndolo<sup>14&#150;16</sup>. La experiencia mundial con este dispositivo ha sido muy buena, hecho que lo ha posicionado como el dispositivo de elecci&oacute;n para la oclusi&oacute;n transcateterismo de la comunicaci&oacute;n interauricular, especialmente de defectos mayores de 20 mm, en los que ning&uacute;n otro dispositivo ha probado su eficacia<sup>14&#150;28</sup>.</font></p>  	    ]]></body>
<body><![CDATA[<p align="justify"><font face="verdana" size="2">En el presente estudio se analiza la experiencia al cumplirse 15 a&ntilde;os del primer caso realizado en nuestro centro, el 24 de septiembre de 1997<sup>18</sup>.</font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>M&eacute;todos</b></font></p>  	    <p align="justify"><font face="verdana" size="2"><b>Pacientes</b></font></p>  	    <p align="justify"><font face="verdana" size="2">El grupo de estudio consta de 806 pacientes consecutivos llevados al laboratorio de hemodinamia para tratamiento percut&aacute;neo de comunicaci&oacute;n interauricular de septiembre de 1997 a diciembre de 2011. De estos 806 pacientes, en 85 no se procedi&oacute; a la colocaci&oacute;n del dispositivo ya que, al realizar el estudio ecocardiogr&aacute;fico en el laboratorio o en el momento de realizar la oclusi&oacute;n con bal&oacute;n del defecto, se consider&oacute; que no eran candidatos a cierre percut&aacute;neo, generalmente por la ausencia de bordes septales en posiciones clave (ausencia de borde a&oacute;rtico y borde posteroinferior) o por considerar el defecto demasiado grande para poder considerar su cierre. De los 721 pacientes restantes, en todos se procedi&oacute; al cierre con la colocaci&oacute;n exitosa inicial del dispositivo (&eacute;xito de colocaci&oacute;n con intenci&oacute;n de tratamiento 89.4&#37;). Este grupo lo componen 216 hombres y 505 mujeres, con edad promedio de 20.6 &#177; 16.5 a&ntilde;os (l&iacute;mites, 1 a 71). En todos los casos se obtuvo el consentimiento informado del paciente o de su representante autorizado (padres).</font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>Procedimiento</b></font></p>  	    <p align="justify"><font face="verdana" size="2">Los pacientes fueron llevados a la sala de cateterismo cardiaco y bajo sedaci&oacute;n profunda o anestesia general e intubaci&oacute;n orotraqueal, con monitorizaci&oacute;n continua del electrocardiograma (ECG) y de la saturaci&oacute;n arterial de ox&iacute;geno transcut&aacute;nea, se realiz&oacute; un ecocardiograma transesof&aacute;gico (HP sonos 1500 con transductor biplanar o multiplanar) y se determin&oacute; la ubicaci&oacute;n y el tama&ntilde;o exactos del defecto en el septum auricular. A partir de julio de 2002, con la introducci&oacute;n en nuestro laboratorio de la ecocardiograf&iacute;a intracardiaca, el procedimiento se realiza bajo sedaci&oacute;n superficial y/o anestesia local y sin necesidad de intubaci&oacute;n. La vena femoral derecha se puncion&oacute; en la forma rutinaria, se llev&oacute; a cabo una evaluaci&oacute;n hemodin&aacute;mica completa y se administraron 50 u/kg de heparina. Para los casos realizados con ecocardiograf&iacute;a intracardiaca, se punciona tambi&eacute;n la vena femoral izquierda, sitio por donde se introduce la sonda Acunav<sup>&#174;</sup> (Acuson, Inc.) con la cual se explora el septum interauricular. Se realiz&oacute; angiograf&iacute;a en la vena pulmonar superior derecha en proyecci&oacute;n oblicua anterior izquierda y con angulaci&oacute;n craneal para demostrar la comunicaci&oacute;n. Se introdujo un cat&eacute;ter con globo (St. Jude Medical) para ocluir el defecto y determinar el &#171;di&aacute;metro expandido&#187;. Se escoge un dispositivo Amplatzer<sup>&#174;</sup> igual, o ligeramente m&aacute;s grande, que el di&aacute;metro expandido del defecto. El introductor femoral se reemplaz&oacute; por un introductor largo con dilatador central (St. Jude Medical) y se introdujo sobre la gu&iacute;a en la aur&iacute;cula izquierda. Se va avanzando el oclusor septal Amplatzer<sup>&#174;</sup> por esta camisa empujando el cable. Bajo control fluorosc&oacute;pico y ecocardiogr&aacute;fico se despliega el disco distal y la cintura central en la porci&oacute;n media de la aur&iacute;cula izquierda y se retira contra el septum auricular. El disco derecho se despliega en la aur&iacute;cula derecha, retrayendo el introductor y manteniendo tracci&oacute;n ligera en el cable. En el 25&#37; de los casos fue necesario realizar alguna maniobra especial para la colocaci&oacute;n adecuada del dispositivo. La posici&oacute;n apropiada del dispositivo a trav&eacute;s de comunicaci&oacute;n interauricular es valorada mediante ecocardiograf&iacute;a. La estabilidad del dispositivo es valorada empujando y arrastrando suavemente con el cable de liberaci&oacute;n. El rotor pl&aacute;stico es conectado al cable de liberaci&oacute;n, el tornillo fijo se aprieta y el rotor se gira en forma antihoraria. La separaci&oacute;n del dispositivo del cable liberador se observa mediante fluoroscopia y por ecocardiograf&iacute;a. Se pasa un cat&eacute;ter angiogr&aacute;fico a la arteria pulmonar o en la aur&iacute;cula derecha y se realiza una angiograf&iacute;a para comprobar la posici&oacute;n del dispositivo y el cierre del defecto. Se finaliza repitiendo las mediciones hemodin&aacute;micas.</font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>Seguimiento</b></font></p>  	    ]]></body>
<body><![CDATA[<p align="justify"><font face="verdana" size="2">Se realiz&oacute; un ECG, radiograf&iacute;a de t&oacute;rax (posteroanterior y oblicua izquierda) y un ecocardiograma bidimensional y doppler de 24 h despu&eacute;s del procedimiento. Si el cierre es satisfactorio y la posici&oacute;n del dispositivo es fija, el paciente es egresado del hospital con un tratamiento de &aacute;cido acetilsalic&iacute;lico de 100 mg/d&iacute;a durante un periodo de 3 a 6 meses y profilaxis para endocarditis bacteriana. Todos los pacientes son valorados en 1, 3, 6, 12 meses y cada a&ntilde;o con examen cl&iacute;nico, ECG, radiograf&iacute;a de t&oacute;rax y/o ecocardiograf&iacute;a.</font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>Estad&iacute;stica</b></font></p>  	    <p align="justify"><font face="verdana" size="2">Los resultados se expresan en media o promedio &#177; una desviaci&oacute;n est&aacute;ndar. La media acotada se utiliz&oacute; en la expresi&oacute;n de los resultados del tiempo del procedimiento y fluoroscopia debido a que algunos de los procedimientos resultaron desproporcionadamente prolongados, ya que se colocaron 2 dispositivos, y para tener una idea exacta del tiempo en el 90&#37; de los procedimientos.</font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>Resultados</b></font></p>  	    <p align="justify"><font face="verdana" size="2">En todos los casos, la auscultaci&oacute;n mostr&oacute; soplo expulsivo pulmonar y desdoblamiento fijo del segundo ruido. En 85 pacientes no hab&iacute;a cardiomegalia en la radiograf&iacute;a y el flujo pulmonar era normal o ligeramente aumentado. Los otros 636 pacientes, con defectos de 12 mm o mayores, ten&iacute;an alg&uacute;n grado de cardiomegalia e hiperflujo. Solo una paciente de 25 a&ntilde;os con defecto de 7 mm ten&iacute;a un ECG normal, el resto presentaban bloqueo incompleto de la rama derecha del haz de His.</font></p>  	    <p align="justify"><font face="verdana" size="2">El defecto fue &uacute;nico en 684 pacientes y se encontraron 2 o m&aacute;s defectos en 37 pacientes (5&#37;), 8 de ellos con fosa oval multifenestrada. El di&aacute;metro m&aacute;ximo del defecto principal medido con ecocardiograf&iacute;a fue de 19 &#177; 6.2 mm (l&iacute;mites, 3 a 37) y el di&aacute;metro expandido fue de 23 &#177; 6.7 mm (l&iacute;mites, 5 a 42). Consideramos defectos peque&ntilde;os aquellos menores de 15 mm de di&aacute;metro expandido, que se presentaron en 85 pacientes (12&#37;); defectos de tama&ntilde;o intermedio entre 15 y 24 mm de di&aacute;metro expandido, con 343 pacientes (48&#37;); y defectos grandes con di&aacute;metro expandido &#8805; 25 mm en 276 pacientes (38&#37;). La presi&oacute;n pulmonar sist&oacute;lica vari&oacute; de 15 a 120 mmHg (promedio, 32.5 &#177; 12.3), con 82 pacientes (11.4&#37;) que ten&iacute;an presi&oacute;n pulmonar sist&oacute;lica &#8805; 50 mmHg. El cociente entre el gasto pulmonar y el gasto sist&eacute;mico promedio fue de 2.32 &#177; 0.87 (l&iacute;mites, 1.1 a 6.8). Hubo 29 pacientes con cociente entre el gasto pulmonar y el gasto sist&eacute;mico mayor de 4, 12 de ellos con cortocircuitos adicionales por defectos asociados (10 con persistencia del conducto arterioso que se cerr&oacute; simult&aacute;neamente y 2 con comunicaci&oacute;n interventricular). Se realiz&oacute; adem&aacute;s valvuloplastia pulmonar en 10 pacientes con estenosis valvular pulmonar y, en 2 casos con estenosis bilateral de ramas pulmonares, se colocaron stents en el mismo procedimiento.</font></p>  	    <p align="justify"><font face="verdana" size="2">De los 37 pacientes con defectos m&uacute;ltiples, estos se lograron ocluir con un solo dispositivo en 33 y con 2 dispositivos que se colocaron simult&aacute;neamente en 4. Una paciente con un defecto secundario de 3 mm que parec&iacute;a haber quedado cubierto con un solo dispositivo, tuvo que ser reintervenida 17 meses despu&eacute;s para colocar un nuevo dispositivo en el defecto residual peque&ntilde;o. En 15 casos (2.1&#37;) fue necesario cambiar el dispositivo porque el colocado inicialmente se consider&oacute; inestable antes de su liberaci&oacute;n; el resto de los dispositivos se consideraron estables y fueron liberadores. En 2 pacientes se inici&oacute; el procedimiento con un dispositivo Helex<sup>&#174;</sup> el cual emboliz&oacute; y se cambi&oacute; por un Amplatzer<sup>&#174;</sup>. El control inmediato con ecocardiograf&iacute;a despu&eacute;s de la liberaci&oacute;n mostr&oacute; oclusi&oacute;n completa de los defectos septales en 247 pacientes (34.2&#37;), fuga m&iacute;nima solo por dentro del dispositivo en 395 (54.8&#37;), fuga ligera (considerada as&iacute; por la presencia de un jet de cortocircuito menor de 1 mm de di&aacute;metro) en 77 (10.7&#37;) y fuga moderada (jet de cortocircuito entre 1 y 2 mm de di&aacute;metro) en 2 pacientes (0.3&#37;). El procedimiento completo en la sala de hemodinamia dur&oacute; entre 10 y 270 min (media acotada, 60) con un tiempo de fluoroscopia entre 2 y 190 min (media acotada, 10.5). Los procedimientos m&aacute;s largos son aquellos en donde se colocaron 2 dispositivos. Se present&oacute; embolizaci&oacute;n del dispositivo dentro de las primeras 24 h despu&eacute;s de su colocaci&oacute;n en 6 pacientes (sin considerar los casos en donde emboliz&oacute; el dispositivo Helex<sup>&#174;</sup>) (0.8&#37;), 4 de ellos en la sala de hemodinamia. En 4 de los casos el dispositivo fue rescatado y recolocado o cambiado por v&iacute;a percut&aacute;nea y solo 2 pacientes, uno con embolizaci&oacute;n a la aur&iacute;cula izquierda y uno con embolizaci&oacute;n al ventr&iacute;culo izquierdo fueron llevados a cirug&iacute;a para la extracci&oacute;n del dispositivo y cierre quir&uacute;rgico del defecto sin complicaciones.</font></p>  	    <p align="justify"><font face="verdana" size="2">La radiograf&iacute;a de t&oacute;rax y el ecocardiograma transtor&aacute;cico realizados antes del alta mostraron adecuada posici&oacute;n de los dispositivos en los 719 pacientes que fueron egresados con dispositivo colocado (&eacute;xito poscolocaci&oacute;n 99.7&#37;, &eacute;xito final con intenci&oacute;n de tratamiento 89.2&#37;).</font></p>  	    ]]></body>
<body><![CDATA[<p align="justify"><font face="verdana" size="2">Se ha logrado seguimiento en 626 pacientes (87.1&#37;) durante un promedio de 33 &#177; 27.5 meses (l&iacute;mites 0.7&#45;150). Durante el periodo de seguimiento se han presentado las siguientes complicaciones: cefalea postoclusi&oacute;n, la cual dura en promedio de 6 a 8 semanas y desaparece en forma espont&aacute;nea, en 54 pacientes (8.6&#37;); 10 pacientes (1.6&#37;) presentaron arritmias supraventriculares nuevas (en 4 de ellos se hab&iacute;an documentado arritmias previas), y en uno de ellos con bloqueo auriculoventricular persistente fue necesaria la colocaci&oacute;n de marcapasos definitivo (0.1&#37;); 4 casos (0.6&#37;) presentaron complicaciones en el sitio de punci&oacute;n, 2 con hematoma inguinal y 2 con f&iacute;stula arteriovenosa femoral; y, finalmente, una paciente (0.1&#37;) que ten&iacute;a fibrilaci&oacute;n auricular previa al cierre del defecto, present&oacute; un accidente cerebrovascular 5 meses despu&eacute;s del cierre del defecto que se presume se present&oacute; porque la paciente suspendi&oacute; el tratamiento anticoagulante despu&eacute;s de ocluido el defecto. El estudio ecocardiogr&aacute;fico mostr&oacute; desaparici&oacute;n del movimiento septal parad&oacute;jico y disminuci&oacute;n del di&aacute;metro telediast&oacute;lico del ventr&iacute;culo derecho de 37 &#177; 6 mm (l&iacute;mites, 18&#45;58) a 26.2 &#177; 4.1 mm (l&iacute;mites, 18&#45;36) (p &#60; 0.00001). En todos los pacientes seguidos excepto en 3, los defectos se encontraron completamente ocluidos despu&eacute;s de 3 meses de seguimiento. De estos, 2 pacientes tienen fuga residual despu&eacute;s de un a&ntilde;o, una enferma, ya comentada previamente, ten&iacute;a 2 defectos que se pens&oacute; inicialmente que quedar&iacute;an cubiertos con un solo dispositivo, pero fue necesario realizar un segundo procedimiento colocando un dispositivo de 6 mm en un defecto residual de 3 mm de di&aacute;metro. El ecocardiograma un mes despu&eacute;s de este &uacute;ltimo procedimiento mostr&oacute; cierre completo de ambos defectos. En nuestra experiencia, no se han presentado embolizaciones de dispositivos despu&eacute;s de las primeras 24 h, ni fracturas de los dispositivos, interferencia con estructuras vecinas o erosiones de estructuras cardiacas como el techo de las aur&iacute;culas o la aorta.</font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>Discusi&oacute;n</b></font></p>  	    <p align="justify"><font face="verdana" size="2">La cirug&iacute;a de la comunicaci&oacute;n interauricular es un m&eacute;todo de tratamiento ya probado, con una baja morbimortalidad que, sin embargo, no est&aacute; exento de complicaciones a largo plazo<sup>1,2</sup>, sin mencionar el tiempo de internamiento, el dolor propio de la herida quir&uacute;rgica y el hecho de tener una cicatriz permanente de esternotom&iacute;a. De hecho, en un estudio publicado recientemente por nuestro grupo, el cierre con dispositivo es superior al tratamiento quir&uacute;rgico en cuanto a la incidencia de complicaciones posprocedimiento, especialmente en pacientes adultos con hipertensi&oacute;n arterial pulmonar<sup>29</sup>. El intento de cierre transcateterismo de estos defectos cardiacos se inici&oacute; en 1976 con King y Mills<sup>3</sup>, y desde entonces, los avances tecnol&oacute;gicos y la creatividad m&eacute;dica han creado diferentes tipos de dispositivos para el cierre percut&aacute;neo de los defectos septales interauriculares<sup>4&#150;13</sup>. La mayor&iacute;a de estos dispositivos constan de 2 discos opuestos que se adosan a cada lado del septum interauricular para cerrar el defecto. El oclusor septal Amplatzer<sup>&#174;</sup> es un dispositivo que tiene caracter&iacute;sticas &uacute;nicas que le proporcionan ventajas sobre los otros dispositivos<sup>15,22,23</sup>. Los resultados obtenidos con su uso lo han colocado como el dispositivo de elecci&oacute;n para el cierre percut&aacute;neo de la comunicaci&oacute;n interauricular, especialmente en defectos mayores de 20 mm, con una alta tasa de &eacute;xito y de cierre completo de los defectos, pr&aacute;cticamente sin complicaciones<sup>14&#150;30</sup>.</font></p>  	    <p align="justify"><font face="verdana" size="2">En nuestra experiencia de 15 a&ntilde;os con el uso de este dispositivo, comprobamos la facilidad de su colocaci&oacute;n con un resultado excelente; de los 721 pacientes en los que se intent&oacute; la colocaci&oacute;n del dispositivo, se logr&oacute; al final cierre completo sin cirug&iacute;a en 719 (99.7&#37;), incluso en aquellos con defectos septales m&uacute;ltiples. Es importante mencionar que, por las caracter&iacute;sticas del dispositivo, no se obtiene oclusi&oacute;n total inmediata del defecto, ya que existe cierto paso de sangre a trav&eacute;s de la malla de alambre y del parche poroso de poli&eacute;ster para promover la trombosis dentro del dispositivo y as&iacute; obtener la oclusi&oacute;n completa y posterior endotelizaci&oacute;n, la cual se ha visto que se completa en 4 a 8 semanas<sup>15</sup>. Esta es la raz&oacute;n de mantener a los pacientes con tratamiento antiagregante plaquetario durante 3 a 6 meses, para evitar fen&oacute;menos emb&oacute;licos, los cuales no se han presentado en los pacientes que han seguido adecuadamente el tratamiento.</font></p>  	    <p align="justify"><font face="verdana" size="2">El uso del ecocardiograma es fundamental en este procedimiento, ya que no solo nos da una visi&oacute;n previa inmejorable del septum interauricular y las mediciones del defecto, sino que es de primordial importancia para el adecuado posicionamiento del dispositivo y la valoraci&oacute;n del resultado inmediato. Desde la introducci&oacute;n de la ecocardiograf&iacute;a intracardiaca en nuestro laboratorio incluso puede realizarse el procedimiento en pacientes adultos, solo con anestesia local y sin necesidad de intubaci&oacute;n<sup>31</sup>.</font></p>  	    <p align="justify"><font face="verdana" size="2">En esta experiencia de 15 a&ntilde;os, con la colocaci&oacute;n en nuestro centro de m&aacute;s de 700 dispositivos, hemos observado que en la mayor&iacute;a de los casos la t&eacute;cnica es r&aacute;pida y sencilla, especialmente en los casos de defectos de tama&ntilde;o peque&ntilde;o o intermedio. El verdadero reto lo constituyen los defectos grandes, especialmente los mayores 25 mm de di&aacute;metro expandido (276 pacientes, 38&#37; de nuestros casos), en los cuales hay que aplicar diversas t&eacute;cnicas como la colocaci&oacute;n desde las venas pulmonares, el apoyo con cat&eacute;ter, bal&oacute;n o la misma sonda de ultrasonido intracardiaco, el uso de camisas de liberaci&oacute;n especial, etc., para poder colocar debidamente el dispositivo en su posici&oacute;n. Estos casos, adem&aacute;s, suele haber ausencia de alguno de los bordes septales, generalmente el anterosuperior o retroa&oacute;rtico, que est&aacute; ausente en m&aacute;s del 70&#37; de los casos con di&aacute;metros mayores de 25 mm, o bien, m&aacute;s raramente, la ausencia del borde septal posteroinferior o de la vena cava inferior, especialmente si se asocia a ausencia del borde a&oacute;rtico, situaciones en las que es preferible no realizar el cierre percut&aacute;neo por la alta incidencia de embolizaci&oacute;n del dispositivo<sup>32,33</sup>.</font></p>  	    <p align="justify"><font face="verdana" size="2">En casos especiales, como en los de levoisomerismo con interrupci&oacute;n de la vena cava inferior intrahep&aacute;tica, se puede realizar el procedimiento sin problema aunque la v&iacute;a de acceso no sea femoral. En estos casos, con acceso femoral ausente, como ocurri&oacute; en 2 de nuestros pacientes, se recomienda el acceso por v&iacute;a transhep&aacute;tica, realizado sin mayores dificultades (<a href="#f1">fig. 1</a>). No mencionaremos las ventajas encontradas con respecto a la cirug&iacute;a, ya que eso ha sido motivo de un estudio publicado previamente<sup>29</sup>; solo recordar que, aun cuando la tasa de morbilidad fuera la misma, el tiempo de estancia hospitalaria es m&aacute;s corto, no se requiere la administraci&oacute;n de hemoderivados, el tiempo de anestesia es menor, y se evita el dolor y la cicatriz de la esternotom&iacute;a.</font></p>  	    <p align="center"><font face="verdana" size="2"><a name="f1"></a></font></p>  	    <p align="center"><font face="verdana" size="2"><img src="/img/revistas/acm/v84n4/a3f1.jpg"></font></p>  	    ]]></body>
<body><![CDATA[<p align="justify"><font face="verdana" size="2">La preocupaci&oacute;n actual con este dispositivo han sido los escasos casos informados de erosi&oacute;n de la ra&iacute;z a&oacute;rtica con f&iacute;stulas hacia las aur&iacute;culas, o del techo de las aur&iacute;culas con el consiguiente taponamiento cardiaco y posible muerte<sup>34,35</sup>. Este hecho aparentemente se ha relacionado con el uso de dispositivos con di&aacute;metro mucho mayor que el tama&ntilde;o del defecto, por lo que se recomienda no sobredimensionar excesivamente los dispositivos<sup>35</sup>. En nuestra experiencia, a&uacute;n no se han presentado casos con esta grave complicaci&oacute;n. El seguimiento ha mostrado la seguridad en nuestra experiencia, con una tasa de complicaciones totales (incluyendo la cefalea autolimitada) del 11&#37; y complicaciones importantes en menos del 1&#37;.</font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>Conclusi&oacute;n</b></font></p>  	    <p align="justify"><font face="verdana" size="2">El dispositivo Amplatzer<sup>&#174;</sup> es un oclusor de comunicaci&oacute;n interauricular cuyo dise&ntilde;o permite el cierre de defectos tipo <i>ostium secundum</i> con una t&eacute;cnica de carga, posicionamiento y liberaci&oacute;n sencilla, en un tiempo relativamente corto y con seguridad para el paciente.</font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>Financiaci&oacute;n</b></font></p>  	    <p align="justify"><font face="verdana" size="2">El presente trabajo no recibi&oacute; financiamiento espec&iacute;fico.</font></p>  	    <p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>Conflicto de intereses</b></font></p>  	    <p align="justify"><font face="verdana" size="2">El Dr. Carlos Zabal es consultor de St. Jude Medical.</font></p>  	    ]]></body>
<body><![CDATA[<p align="justify"><font face="verdana" size="2">&nbsp;</font></p>  	    <p align="justify"><font face="verdana" size="2"><b>Bibliograf&iacute;a</b></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">1. D. Young. Later results of closure of secundum atrial septal defect in children. Am J Cardiol. 1973;31:14&#45;22.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132225&pid=S1405-9940201400040000300001&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">2. J.G. Murphy, B.J. Gersh, M.D. McGoon. Long&#45;term outcome after surgical repair of isolated atrial septal defect. N Engl J Med. 1990;323:1645&#45;50.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132227&pid=S1405-9940201400040000300002&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">3. T.D. King, N.L. Mills. Secundum atrial septal defect: Nonoperative closure during cardiac catheterization. JAMA. 1976;235:2506&#45;9.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132229&pid=S1405-9940201400040000300003&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">4. E.B. Sideris, S.E. Sideris, B.D. Thanapoulos. Transvenous atrial septal defect occlusion by the buttoned device. Am J Cardiol. 1990;66:1524&#45;6.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132231&pid=S1405-9940201400040000300004&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    ]]></body>
<body><![CDATA[<!-- ref --><p align="justify"><font face="verdana" size="2">5. T.R. Lloyd, P.S. Rao, R.H. Beekmam III. Atrial septal defect occlusion with the buttoned device (a multi&#45;institutional U.S. trial). Am J Cardiol. 1994;73:286&#45;91.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132233&pid=S1405-9940201400040000300005&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">6. P.S. Rao, E.B. Sideris, G. Hausdorf. International experience with secundum atrial septal defect occlusion by the buttoned device. Am Heart J. 1994;128:1022&#45;35.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132235&pid=S1405-9940201400040000300006&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">7. H. Sievert, U.U. Babic, R. Ensslen. Transcatheter closure of large atrial septal defects with the Babic system. Cath Cardiovasc Diag. 1995;36:232&#45;40.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132237&pid=S1405-9940201400040000300007&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">8. G. Hausdorf, M. Schneider, B. Franzbach. Transcatheter closure of secundum atrial septal defects with the atrial septal defect occlusion system (ASDOS): Initial experience in children. Heart. 1996;75:83&#45;8.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132239&pid=S1405-9940201400040000300008&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">9. J.E. Lock, J.J. Rome, R. Davis. Transcatheter closure of atrial septal defects. Experimental studies. Circulation. 1989;79:1091&#45;9.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132241&pid=S1405-9940201400040000300009&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    ]]></body>
<body><![CDATA[<!-- ref --><p align="justify"><font face="verdana" size="2">10. J.J. Rome, J.F. Keane, S.B. Perry. Double&#45;umbrella closure of atrial defects. Initial clinical applications. Circulation. 1990;82:751&#45;8.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132243&pid=S1405-9940201400040000300010&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">11. D. Pavenik, K.C. Wright, S. Wallace. Monodisk device for percutaneous transcatheter closure of cardiac septal defects. Cardiovasc Intervent Radiol. 1993;16:308&#45;12.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132245&pid=S1405-9940201400040000300011&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">12. G.S. Das, G. Voss, G. Jarvis. Experimental atrial septal defect closure with a new transcatheter, self&#45;centering device. Circulation. 1993;88:1754&#45;64.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132247&pid=S1405-9940201400040000300012&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">13. G.S. Das, Z.M. Hijazi, M.P. O'Laughlin. Initial results of the US PFO/ASD closure trial. Circulation. 1996;27:119.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132249&pid=S1405-9940201400040000300013&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">14. P.G. Bj&#248;rnstad, B. Smevik, A.E. Fiane. Catheter&#45;based closure of atrial septal defects with a newly developed nitinol double disc: An experimental study. Cardiol Young. 1997;7:220&#45;4.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132251&pid=S1405-9940201400040000300014&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    ]]></body>
<body><![CDATA[<!-- ref --><p align="justify"><font face="verdana" size="2">15. M.J.A. Sharafuddin, X. Gu, J.L. Titus. Transvenous closure of secundum atrial septal defects. Preliminary results with a new self&#45;expanding Nitinol prosthesis in a swine model. Circulation. 1997;95:2162&#45;8.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132253&pid=S1405-9940201400040000300015&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">16. P.G. Bj&#248;rnstad, J. Masura, E. Thaulow. Interventional closure of atrial septal defects with the Amplatzer<sup>&#174;</sup> device: First clinical experience. Cardiol Young. 1997;7:277&#45;83.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132255&pid=S1405-9940201400040000300016&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">17. F. Hakim, A. Madani, Y. Samara. Transcatheter closure of secundum atrial septal defect in a patient with dextrocardia using the Amplatzer<sup>&#174;</sup> septal occluder. Cathet Cardiovasc Diagn. 1998;43:291&#45;4.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132257&pid=S1405-9940201400040000300017&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">18. C. Zabal, F. Chio, K. Amplatz. Cierre percut&aacute;neo de comunicacion interatial con el dispositivo Amplatzer<sup>&#174;</sup>. Primer caso en M&eacute;xico. Arch Inst Cardiol Mex. 1998;68:147&#45;52.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132259&pid=S1405-9940201400040000300018&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">19. J.L. Wilkinson, T.H. Goh. Early clinical experience with use of Amplatzer<sup>&#174;</sup> septal occluder device for atrial septal defect. Cardiol Young. 1998;8:295&#45;302.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132261&pid=S1405-9940201400040000300019&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    ]]></body>
<body><![CDATA[<!-- ref --><p align="justify"><font face="verdana" size="2">20. P.G. Bj&#248;rnstad. The role of devices in the closure of atrial septal defects in the oval fossa. Cardiol Young. 1998;8:285&#45;6.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132263&pid=S1405-9940201400040000300020&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">21. R. Formigari, G. Santoro, L. Rossetti. Comparison of three different atrial septal defect occlusion devices. Am J Cardiol. 1998;82:690&#45;2.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132265&pid=S1405-9940201400040000300021&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">22. K.Y. Chan, W.C. Yip, M.J. Godman. Transcatheter occlusion of atrial septal defects: An initial experience with the Amplatzer<sup>&#174;</sup> septal occluder. J Paediatr Child Health. 1998;34:369&#45;73.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132267&pid=S1405-9940201400040000300022&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">23. V.F. Fontes, C.A. Pedra, S.R. Pedra. Initial experience in percutaneous closure of interatrial communication with the Amplatzer<sup>&#174;</sup> device. Arq Bras Cardiol. 1998;70:147&#45;53.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132269&pid=S1405-9940201400040000300023&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">24. F. Berger, P. Ewert, B. Stiller. Initial clinical results with the Amplatzer<sup>&#174;</sup> septal occluder&#45; a self&#45;centering double disc for occlusion of atrial septal defects. Z Kardiol. 1998;87:185&#45;90.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132271&pid=S1405-9940201400040000300024&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    ]]></body>
<body><![CDATA[<!-- ref --><p align="justify"><font face="verdana" size="2">25. F. Berger, M. Vogel, V. Alexi&#45;Meskishvili. Comparison of results and complications of surgical and Amplatzer device closure of atrial septal defects. J Thorac Cardiovasc Surg. 1999;118:674&#45;8.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132273&pid=S1405-9940201400040000300025&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">26. P. Ewert, F. Berger, O. Kretschmar. Feasibility of transcatheter closure of multiple defects within the oval fossa. Cardiol Young. 2001;11:314&#45;9.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132275&pid=S1405-9940201400040000300026&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">27. A. Omeish, Z.M. Hijazi. Transcatheter closure of atrial septal defects in children and adults using the Amplatzer septal occluder. J Interv Cardiol. 2001;14:37&#45;44.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132277&pid=S1405-9940201400040000300027&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">28. F. Berger, P. Ewert, H. Abdul&#45;Khaliq. Percutaneous closure of large atrial septal defects with the Amplatzer septal occluder: Technical overkill or recommendable alternative treatment. J Interv Cardiol. 2001;14:63&#45;7.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132279&pid=S1405-9940201400040000300028&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">29. M. Rosas, C. Zabal, J. Garc&iacute;a&#45;Montes. Transcatheter versus surgical closure of secundum atrial septal defect in adults: Impact of age at intervention. A concurrent matched comparative study. Congenit Heart Dis. 2007;2:148&#45;55.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132281&pid=S1405-9940201400040000300029&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    ]]></body>
<body><![CDATA[<!-- ref --><p align="justify"><font face="verdana" size="2">30. G. Webb, M.A. Gatzoulis. Atrial septal defect in the adult: Recent progress and overview. Circulation. 2006;114:1645&#45;53.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132283&pid=S1405-9940201400040000300030&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">31. C. Zabal. Utilidad de la ecocardiograf&iacute;a intracard&iacute;aca en el cateterismo intervencionista de las cardiopat&iacute;as cong&eacute;nitas. Arch Cardiol Mex. 2004;74:S113&#45;8.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132285&pid=S1405-9940201400040000300031&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">32. B.R. Kannan, E. Francis, K. Sivakumar. Transcatheter closure of very large (&#62;or&#61; 25 mm) atrial septal defects using the Amplatzer septal occluder. Catheter Cardiovasc Interv. 2003;59:522&#45;7.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132287&pid=S1405-9940201400040000300032&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">33. C. Varma, L.N. Benson, C. Silversides. Outcomes and alternative techniques for device closure of the large secundum atrial septal defect. Catheter Cardiovasc Interv. 2004;61:131&#45;9.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132289&pid=S1405-9940201400040000300033&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    <!-- ref --><p align="justify"><font face="verdana" size="2">34. P.A. Grayburn, B. Schwartz, A. Anwar. Migration of an Amplatzer septal occluder device for closure of atrial septal defect into the ascending aorta with formation of an aorta&#45;to&#45;right atrial fistula. Am J Cardiol. 2005;96:1607&#45;9.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132291&pid=S1405-9940201400040000300034&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>  	    ]]></body>
<body><![CDATA[<!-- ref --><p align="justify"><font face="verdana" size="2">35. Z. Amin, Z.M. Hijazi, J.L. Bass. Erosion of Amplatzer septal occluder device after closure of secundum atrial septal defects: Review of registry of complications and recommendations to minimize future risk. Catheter Cardiovasc Interv. 2004;63:496&#45;502.    &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;[&#160;<a href="javascript:void(0);" onclick="javascript: window.open('/scielo.php?script=sci_nlinks&ref=1132293&pid=S1405-9940201400040000300035&lng=','','width=640,height=500,resizable=yes,scrollbars=1,menubar=yes,');">Links</a>&#160;]<!-- end-ref --></font></p>      ]]></body><back>
<ref-list>
<ref id="B1">
<label>1</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Young]]></surname>
<given-names><![CDATA[D.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Later results of closure of secundum atrial septal defect in children]]></article-title>
<source><![CDATA[Am J Cardiol]]></source>
<year>1973</year>
<volume>31</volume>
<page-range>14-22</page-range></nlm-citation>
</ref>
<ref id="B2">
<label>2</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Murphy]]></surname>
<given-names><![CDATA[J.G.]]></given-names>
</name>
<name>
<surname><![CDATA[Gersh]]></surname>
<given-names><![CDATA[B.J.]]></given-names>
</name>
<name>
<surname><![CDATA[McGoon]]></surname>
<given-names><![CDATA[M.D.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Long-term outcome after surgical repair of isolated atrial septal defect]]></article-title>
<source><![CDATA[N Engl J Med]]></source>
<year>1990</year>
<volume>323</volume>
<page-range>1645-50</page-range></nlm-citation>
</ref>
<ref id="B3">
<label>3</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[King]]></surname>
<given-names><![CDATA[T.D.]]></given-names>
</name>
<name>
<surname><![CDATA[Mills]]></surname>
<given-names><![CDATA[N.L.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Secundum atrial septal defect: Nonoperative closure during cardiac catheterization]]></article-title>
<source><![CDATA[JAMA]]></source>
<year>1976</year>
<volume>235</volume>
<page-range>2506-9</page-range></nlm-citation>
</ref>
<ref id="B4">
<label>4</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Sideris]]></surname>
<given-names><![CDATA[E.B.]]></given-names>
</name>
<name>
<surname><![CDATA[Sideris]]></surname>
<given-names><![CDATA[S.E.]]></given-names>
</name>
<name>
<surname><![CDATA[Thanapoulos]]></surname>
<given-names><![CDATA[B.D.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Transvenous atrial septal defect occlusion by the buttoned device]]></article-title>
<source><![CDATA[Am J Cardiol]]></source>
<year>1990</year>
<volume>66</volume>
<page-range>1524-6</page-range></nlm-citation>
</ref>
<ref id="B5">
<label>5</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Lloyd]]></surname>
<given-names><![CDATA[T.R.]]></given-names>
</name>
<name>
<surname><![CDATA[Rao]]></surname>
<given-names><![CDATA[P.S.]]></given-names>
</name>
<name>
<surname><![CDATA[Beekmam III]]></surname>
<given-names><![CDATA[R.H.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Atrial septal defect occlusion with the buttoned device (a multi-institutional U.S. trial)]]></article-title>
<source><![CDATA[Am J Cardiol]]></source>
<year>1994</year>
<volume>73</volume>
<page-range>286-91</page-range></nlm-citation>
</ref>
<ref id="B6">
<label>6</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Rao]]></surname>
<given-names><![CDATA[P.S.]]></given-names>
</name>
<name>
<surname><![CDATA[Sideris]]></surname>
<given-names><![CDATA[E.B.]]></given-names>
</name>
<name>
<surname><![CDATA[Hausdorf]]></surname>
<given-names><![CDATA[G.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[International experience with secundum atrial septal defect occlusion by the buttoned device]]></article-title>
<source><![CDATA[Am Heart J]]></source>
<year>1994</year>
<volume>128</volume>
<page-range>1022-35</page-range></nlm-citation>
</ref>
<ref id="B7">
<label>7</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Sievert]]></surname>
<given-names><![CDATA[H.]]></given-names>
</name>
<name>
<surname><![CDATA[Babic]]></surname>
<given-names><![CDATA[U.U.]]></given-names>
</name>
<name>
<surname><![CDATA[Ensslen]]></surname>
<given-names><![CDATA[R.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Transcatheter closure of large atrial septal defects with the Babic system]]></article-title>
<source><![CDATA[Cath Cardiovasc Diag]]></source>
<year>1995</year>
<volume>36</volume>
<page-range>232-40</page-range></nlm-citation>
</ref>
<ref id="B8">
<label>8</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Hausdorf]]></surname>
<given-names><![CDATA[G.]]></given-names>
</name>
<name>
<surname><![CDATA[Schneider]]></surname>
<given-names><![CDATA[M.]]></given-names>
</name>
<name>
<surname><![CDATA[Franzbach]]></surname>
<given-names><![CDATA[B.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Transcatheter closure of secundum atrial septal defects with the atrial septal defect occlusion system (ASDOS): Initial experience in children]]></article-title>
<source><![CDATA[Heart]]></source>
<year>1996</year>
<volume>75</volume>
<page-range>83-8</page-range></nlm-citation>
</ref>
<ref id="B9">
<label>9</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Lock]]></surname>
<given-names><![CDATA[J.E.]]></given-names>
</name>
<name>
<surname><![CDATA[Rome]]></surname>
<given-names><![CDATA[J.J.]]></given-names>
</name>
<name>
<surname><![CDATA[Davis]]></surname>
<given-names><![CDATA[R.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Transcatheter closure of atrial septal defects. Experimental studies]]></article-title>
<source><![CDATA[Circulation]]></source>
<year>1989</year>
<volume>79</volume>
<page-range>1091-9</page-range></nlm-citation>
</ref>
<ref id="B10">
<label>10</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Rome]]></surname>
<given-names><![CDATA[J.J.]]></given-names>
</name>
<name>
<surname><![CDATA[Keane]]></surname>
<given-names><![CDATA[J.F.]]></given-names>
</name>
<name>
<surname><![CDATA[Perry]]></surname>
<given-names><![CDATA[S.B.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Double-umbrella closure of atrial defects. Initial clinical applications]]></article-title>
<source><![CDATA[Circulation]]></source>
<year>1990</year>
<volume>82</volume>
<page-range>751-8</page-range></nlm-citation>
</ref>
<ref id="B11">
<label>11</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[D.]]></surname>
<given-names><![CDATA[Pavenik]]></given-names>
</name>
<name>
<surname><![CDATA[Wright]]></surname>
<given-names><![CDATA[K.C.]]></given-names>
</name>
<name>
<surname><![CDATA[Wallace]]></surname>
<given-names><![CDATA[S.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Monodisk device for percutaneous transcatheter closure of cardiac septal defects]]></article-title>
<source><![CDATA[Cardiovasc Intervent Radiol]]></source>
<year>1993</year>
<volume>16</volume>
<page-range>308-12</page-range></nlm-citation>
</ref>
<ref id="B12">
<label>12</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Das]]></surname>
<given-names><![CDATA[G.S.]]></given-names>
</name>
<name>
<surname><![CDATA[Voss]]></surname>
<given-names><![CDATA[G.]]></given-names>
</name>
<name>
<surname><![CDATA[Jarvis]]></surname>
<given-names><![CDATA[G.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Experimental atrial septal defect closure with a new transcatheter, self-centering device]]></article-title>
<source><![CDATA[Circulation]]></source>
<year>1993</year>
<volume>88</volume>
<page-range>1754-64</page-range></nlm-citation>
</ref>
<ref id="B13">
<label>13</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Das]]></surname>
<given-names><![CDATA[G.S.]]></given-names>
</name>
<name>
<surname><![CDATA[Hijazi]]></surname>
<given-names><![CDATA[Z.M.]]></given-names>
</name>
<name>
<surname><![CDATA[O'Laughlin]]></surname>
<given-names><![CDATA[M.P.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Initial results of the US PFO/ASD closure trial]]></article-title>
<source><![CDATA[Circulation]]></source>
<year>1996</year>
<volume>27</volume>
<page-range>119</page-range></nlm-citation>
</ref>
<ref id="B14">
<label>14</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Bjørnstad]]></surname>
<given-names><![CDATA[P.G.]]></given-names>
</name>
<name>
<surname><![CDATA[Smevik]]></surname>
<given-names><![CDATA[B.]]></given-names>
</name>
<name>
<surname><![CDATA[Fiane]]></surname>
<given-names><![CDATA[A.E.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Catheter-based closure of atrial septal defects with a newly developed nitinol double disc: An experimental study]]></article-title>
<source><![CDATA[Cardiol Young]]></source>
<year>1997</year>
<volume>7</volume>
<page-range>220-4</page-range></nlm-citation>
</ref>
<ref id="B15">
<label>15</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Sharafuddin]]></surname>
<given-names><![CDATA[M.J.A.]]></given-names>
</name>
<name>
<surname><![CDATA[Gu]]></surname>
<given-names><![CDATA[X.]]></given-names>
</name>
<name>
<surname><![CDATA[Titus]]></surname>
<given-names><![CDATA[J.L.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Transvenous closure of secundum atrial septal defects. Preliminary results with a new self-expanding Nitinol prosthesis in a swine model]]></article-title>
<source><![CDATA[Circulation]]></source>
<year>1997</year>
<volume>95</volume>
<page-range>2162-8</page-range></nlm-citation>
</ref>
<ref id="B16">
<label>16</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Bjørnstad]]></surname>
<given-names><![CDATA[P.G.]]></given-names>
</name>
<name>
<surname><![CDATA[Masura]]></surname>
<given-names><![CDATA[J.]]></given-names>
</name>
<name>
<surname><![CDATA[Thaulow]]></surname>
<given-names><![CDATA[E.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Interventional closure of atrial septal defects with the Amplatzer® device: First clinical experience]]></article-title>
<source><![CDATA[Cardiol Young]]></source>
<year>1997</year>
<volume>7</volume>
<page-range>277-83</page-range></nlm-citation>
</ref>
<ref id="B17">
<label>17</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Hakim]]></surname>
<given-names><![CDATA[F.]]></given-names>
</name>
<name>
<surname><![CDATA[Madani]]></surname>
<given-names><![CDATA[A.]]></given-names>
</name>
<name>
<surname><![CDATA[Samara]]></surname>
<given-names><![CDATA[Y.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Transcatheter closure of secundum atrial septal defect in a patient with dextrocardia using the Amplatzer® septal occluder]]></article-title>
<source><![CDATA[Cathet Cardiovasc Diagn]]></source>
<year>1998</year>
<volume>43</volume>
<page-range>291-4</page-range></nlm-citation>
</ref>
<ref id="B18">
<label>18</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Zabal]]></surname>
<given-names><![CDATA[C.]]></given-names>
</name>
<name>
<surname><![CDATA[Chio]]></surname>
<given-names><![CDATA[F.]]></given-names>
</name>
<name>
<surname><![CDATA[Amplatz]]></surname>
<given-names><![CDATA[K.]]></given-names>
</name>
</person-group>
<article-title xml:lang="es"><![CDATA[Cierre percutáneo de comunicacion interatial con el dispositivo Amplatzer®. Primer caso en México]]></article-title>
<source><![CDATA[Arch Inst Cardiol Mex]]></source>
<year>1998</year>
<volume>68</volume>
<page-range>147-52</page-range></nlm-citation>
</ref>
<ref id="B19">
<label>19</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Wilkinson]]></surname>
<given-names><![CDATA[J.L.]]></given-names>
</name>
<name>
<surname><![CDATA[Goh]]></surname>
<given-names><![CDATA[T.H.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Early clinical experience with use of Amplatzer® septal occluder device for atrial septal defect]]></article-title>
<source><![CDATA[Cardiol Young]]></source>
<year>1998</year>
<volume>8</volume>
<page-range>295-302</page-range></nlm-citation>
</ref>
<ref id="B20">
<label>20</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Bjørnstad]]></surname>
<given-names><![CDATA[P.G.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[The role of devices in the closure of atrial septal defects in the oval fossa]]></article-title>
<source><![CDATA[Cardiol Young]]></source>
<year>1998</year>
<volume>8</volume>
<page-range>285-6</page-range></nlm-citation>
</ref>
<ref id="B21">
<label>21</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Formigari]]></surname>
<given-names><![CDATA[R.]]></given-names>
</name>
<name>
<surname><![CDATA[Santoro]]></surname>
<given-names><![CDATA[G.]]></given-names>
</name>
<name>
<surname><![CDATA[Rossetti]]></surname>
<given-names><![CDATA[L.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Comparison of three different atrial septal defect occlusion devices]]></article-title>
<source><![CDATA[Am J Cardiol]]></source>
<year>1998</year>
<volume>82</volume>
<page-range>690-2</page-range></nlm-citation>
</ref>
<ref id="B22">
<label>22</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Chan]]></surname>
<given-names><![CDATA[K.Y.]]></given-names>
</name>
<name>
<surname><![CDATA[Yip]]></surname>
<given-names><![CDATA[W.C.]]></given-names>
</name>
<name>
<surname><![CDATA[Godman]]></surname>
<given-names><![CDATA[M.J.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Transcatheter occlusion of atrial septal defects: An initial experience with the Amplatzer® septal occluder]]></article-title>
<source><![CDATA[J Paediatr Child Health]]></source>
<year>1998</year>
<volume>34</volume>
<page-range>369-73</page-range></nlm-citation>
</ref>
<ref id="B23">
<label>23</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Fontes]]></surname>
<given-names><![CDATA[V.F.]]></given-names>
</name>
<name>
<surname><![CDATA[Pedra]]></surname>
<given-names><![CDATA[C.A.]]></given-names>
</name>
<name>
<surname><![CDATA[Pedra]]></surname>
<given-names><![CDATA[S.R.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Initial experience in percutaneous closure of interatrial communication with the Amplatzer® device]]></article-title>
<source><![CDATA[Arq Bras Cardiol]]></source>
<year>1998</year>
<volume>70</volume>
<page-range>147-53</page-range></nlm-citation>
</ref>
<ref id="B24">
<label>24</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Berger]]></surname>
<given-names><![CDATA[F.]]></given-names>
</name>
<name>
<surname><![CDATA[Ewert]]></surname>
<given-names><![CDATA[P.]]></given-names>
</name>
<name>
<surname><![CDATA[Stiller]]></surname>
<given-names><![CDATA[B.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Initial clinical results with the Amplatzer® septal occluder- a self-centering double disc for occlusion of atrial septal defects]]></article-title>
<source><![CDATA[Z Kardiol]]></source>
<year>1998</year>
<volume>87</volume>
<page-range>185-90</page-range></nlm-citation>
</ref>
<ref id="B25">
<label>25</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Berger]]></surname>
<given-names><![CDATA[F.]]></given-names>
</name>
<name>
<surname><![CDATA[Vogel]]></surname>
<given-names><![CDATA[M.]]></given-names>
</name>
<name>
<surname><![CDATA[Alexi-Meskishvili]]></surname>
<given-names><![CDATA[V.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Comparison of results and complications of surgical and Amplatzer device closure of atrial septal defects]]></article-title>
<source><![CDATA[J Thorac Cardiovasc Surg]]></source>
<year>1999</year>
<volume>118</volume>
<page-range>674-8</page-range></nlm-citation>
</ref>
<ref id="B26">
<label>26</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Ewert]]></surname>
<given-names><![CDATA[P.]]></given-names>
</name>
<name>
<surname><![CDATA[Berger]]></surname>
<given-names><![CDATA[F.]]></given-names>
</name>
<name>
<surname><![CDATA[Kretschmar]]></surname>
<given-names><![CDATA[O.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Feasibility of transcatheter closure of multiple defects within the oval fossa]]></article-title>
<source><![CDATA[Cardiol Young]]></source>
<year>2001</year>
<volume>11</volume>
<page-range>314-9</page-range></nlm-citation>
</ref>
<ref id="B27">
<label>27</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Omeish]]></surname>
<given-names><![CDATA[A.]]></given-names>
</name>
<name>
<surname><![CDATA[Hijazi]]></surname>
<given-names><![CDATA[Z.M.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Transcatheter closure of atrial septal defects in children and adults using the Amplatzer septal occluder]]></article-title>
<source><![CDATA[J Interv Cardiol]]></source>
<year>2001</year>
<volume>14</volume>
<page-range>37-44</page-range></nlm-citation>
</ref>
<ref id="B28">
<label>28</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Berger]]></surname>
<given-names><![CDATA[F.]]></given-names>
</name>
<name>
<surname><![CDATA[Ewert]]></surname>
<given-names><![CDATA[P.]]></given-names>
</name>
<name>
<surname><![CDATA[Abdul-Khaliq]]></surname>
<given-names><![CDATA[H.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Percutaneous closure of large atrial septal defects with the Amplatzer septal occluder: Technical overkill or recommendable alternative treatment]]></article-title>
<source><![CDATA[J Interv Cardiol]]></source>
<year>2001</year>
<volume>14</volume>
<page-range>63-7</page-range></nlm-citation>
</ref>
<ref id="B29">
<label>29</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Rosas]]></surname>
<given-names><![CDATA[M.]]></given-names>
</name>
<name>
<surname><![CDATA[Zabal]]></surname>
<given-names><![CDATA[C.]]></given-names>
</name>
<name>
<surname><![CDATA[García-Montes]]></surname>
<given-names><![CDATA[J.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Transcatheter versus surgical closure of secundum atrial septal defect in adults: Impact of age at intervention. A concurrent matched comparative study]]></article-title>
<source><![CDATA[Congenit Heart Dis]]></source>
<year>2007</year>
<volume>2</volume>
<page-range>148-55</page-range></nlm-citation>
</ref>
<ref id="B30">
<label>30</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Webb]]></surname>
<given-names><![CDATA[G.]]></given-names>
</name>
<name>
<surname><![CDATA[Gatzoulis]]></surname>
<given-names><![CDATA[M.A.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Atrial septal defect in the adult: Recent progress and overview]]></article-title>
<source><![CDATA[Circulation]]></source>
<year>2006</year>
<volume>114</volume>
<page-range>1645-53</page-range></nlm-citation>
</ref>
<ref id="B31">
<label>31</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Zabal]]></surname>
<given-names><![CDATA[C.]]></given-names>
</name>
</person-group>
<article-title xml:lang="es"><![CDATA[Utilidad de la ecocardiografía intracardíaca en el cateterismo intervencionista de las cardiopatías congénitas]]></article-title>
<source><![CDATA[Arch Cardiol Mex]]></source>
<year>2004</year>
<volume>74</volume>
<page-range>S113-8</page-range></nlm-citation>
</ref>
<ref id="B32">
<label>32</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Kannan]]></surname>
<given-names><![CDATA[B.R.]]></given-names>
</name>
<name>
<surname><![CDATA[Francis]]></surname>
<given-names><![CDATA[E.]]></given-names>
</name>
<name>
<surname><![CDATA[Sivakumar]]></surname>
<given-names><![CDATA[K.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Transcatheter closure of very large (>or= 25 mm) atrial septal defects using the Amplatzer septal occluder]]></article-title>
<source><![CDATA[Catheter Cardiovasc Interv]]></source>
<year>2003</year>
<volume>59</volume>
<page-range>522-7</page-range></nlm-citation>
</ref>
<ref id="B33">
<label>33</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Varma]]></surname>
<given-names><![CDATA[C.]]></given-names>
</name>
<name>
<surname><![CDATA[Benson]]></surname>
<given-names><![CDATA[L.N.]]></given-names>
</name>
<name>
<surname><![CDATA[Silversides]]></surname>
<given-names><![CDATA[C.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Outcomes and alternative techniques for device closure of the large secundum atrial septal defect]]></article-title>
<source><![CDATA[Catheter Cardiovasc Interv]]></source>
<year>2004</year>
<volume>61</volume>
<page-range>131-9</page-range></nlm-citation>
</ref>
<ref id="B34">
<label>34</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Grayburn]]></surname>
<given-names><![CDATA[P.A.]]></given-names>
</name>
<name>
<surname><![CDATA[Schwartz]]></surname>
<given-names><![CDATA[B.]]></given-names>
</name>
<name>
<surname><![CDATA[Anwar]]></surname>
<given-names><![CDATA[A.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Migration of an Amplatzer septal occluder device for closure of atrial septal defect into the ascending aorta with formation of an aorta-to-right atrial fistula]]></article-title>
<source><![CDATA[Am J Cardiol]]></source>
<year>2005</year>
<volume>96</volume>
<page-range>1607-9</page-range></nlm-citation>
</ref>
<ref id="B35">
<label>35</label><nlm-citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname><![CDATA[Amin]]></surname>
<given-names><![CDATA[Z.]]></given-names>
</name>
<name>
<surname><![CDATA[Hijazi]]></surname>
<given-names><![CDATA[Z.M.]]></given-names>
</name>
<name>
<surname><![CDATA[Bass]]></surname>
<given-names><![CDATA[J.L.]]></given-names>
</name>
</person-group>
<article-title xml:lang="en"><![CDATA[Erosion of Amplatzer septal occluder device after closure of secundum atrial septal defects: Review of registry of complications and recommendations to minimize future risk]]></article-title>
<source><![CDATA[Catheter Cardiovasc Interv]]></source>
<year>2004</year>
<volume>63</volume>
<page-range>496-502</page-range></nlm-citation>
</ref>
</ref-list>
</back>
</article>