Serviços Personalizados
Journal
Artigo
Indicadores
- Citado por SciELO
- Acessos
Links relacionados
- Similares em SciELO
Compartilhar
Superficies y vacío
versão impressa ISSN 1665-3521
Resumo
PEREZ-LUNA, J. G. et al. Técnica de medición de baja corriente para un convertidor termoiónico. Superf. vacío [online]. 2014, vol.27, n.1, pp.30-34. ISSN 1665-3521.
An analog electrical technique is presented to measure a small current that is generated by a thermionic converter when it is operated at low temperature regime. Direct and inverse currents are generated simultaneously in a thermionic converter at steady state (emitter temperature > 1500K); although, at low temperature regime any of them may be more significant. Important information is obtained from a thermionic converter when it is operated at low temperature regime, particularly, when the inverse current is generated from the collector to the emitter (inverse current ≈ 1 µA). An operational amplifier is used as a buffer circuit to measure the voltage in a small capacitor. A mathematical treatment was made from the voltage data obtained from the charged capacitor to measure indirectly the current generated by the thermionic converter. It was possible to measure a small current like 1 µA by this technique.
Palavras-chave : Measurement technique; Small current; Thermionic converter.