vol.79 issue2Clinical interpretation for the pressure-flow relationships in extrinsic allergic alveolitis and in interstitial lung disease pulmonary hypertension patients: Should we care for the lung, the pulmonary artery pressure or both?Truncus arteriosus, anatomy, surgical treatment, follow up author indexsubject indexsearch form
Home Pagealphabetic serial listing  

Services on Demand

Journal

Article

Indicators

Related links

  • Have no similar articlesSimilars in SciELO

Share


Archivos de cardiología de México

On-line version ISSN 1665-1731Print version ISSN 1405-9940

Abstract

MUNAYER-CALDERON, Jaime E et al. Percutaneous closure of inter-atrial communication with the Amplatzer device: Experience with 42 cases. Arch. Cardiol. Méx. [online]. 2009, vol.79, n.2, pp.104-106. ISSN 1665-1731.

We report our experience on 42 patients treated with atrial septal defect (ASD) occlusion using Amplatzer device. Thirty patients were females and 12 were males, mean ages 26.5 years-old ± 12 years (interval from 7 to 69 years). Average weight was 57.1 ± 13.8 kg (32.3-85.0 kg). Forty-two devices (ranging diameters from 13 mm to 40 mm) were deployed, 40 of which were placed successfully and attempts were unsuccessful in two cases (diameter devices 36 mm and 40 mm, respectively). We used the "balloon on the left or right upper pulmonary vein" in 5 patients, achieving good deployment. Echocardiography showed total occlusion in 37 patients (93.5%), trivial leak in 2 (4.7%), and light leak in 1 patient (2.3%). Follow up was at 1 to 12 months (mean 6.5). Total occlusion was observed at one month on both patients with trivial leak, and at 6 months on the patient with light leak. Failure to deploy the device appropriately on the two patients with unsuccessful result was due to unfavorable anatomy: very large defects in both cases (occluder size 36 mm and 40 mm), very thin postero-superior 6 mm rim on one of them and aortic rim absence on the other one. Stretched diameters were 34 mm and 38 mm on patients with 36 mm and 40 mm devices, respectively. Both of them were sent to surgery. We conclude that percutaneus closure of atrial septal defect with the Amplatzer device is a save and have good results.

Keywords : Amplatzer device; Echocardiogram; Atrial septal defect; Mexico.