Services on Demand
Journal
Article
Indicators
- Cited by SciELO
- Access statistics
Related links
- Similars in SciELO
Share
Revista mexicana de física
Print version ISSN 0035-001X
Abstract
RODRIGUEZ-VILLAFUERTE, M.; ALVA-SANCHEZ, H.; MURRIETA, T. and MARTINEZ-DAVALOS, A. Espectroscopía gamma con LaBr3 y LYSO para su aplicación en medicina nuclear. Rev. mex. fis. [online]. 2011, vol.57, n.1, pp.91-97. ISSN 0035-001X.
Image formation in nuclear medicine is based on the detection of radiation emitted from a small quantity of a radiopharmaceutical administered to patients. This requires high detection efficiency, good spatial resolution and the ability to identify the energy of the incident radiation. These and other properties of detector materials for gamma radiation are reviewed. Standard detector calibration and characterization procedures are described, and the results from gamma spectroscopy measurements for sodium iodide (NaI:Tl), conventional scintillator widely used in nuclear medicine applications, and two other novel scintillation crystals, lanthanum bromide (LaBr3) and lutetium-yttrium oxyorthosilicate (LYSO) are presented. Both materials showed a linear energy response from 80 to 1332 keV, and within this energy interval, LaBr3 had a superior energy resolution compared to LYSO and NaI:Tl. These and other new materials are expected to replace those used in current detectors for nuclear medicine scanners.
Keywords : Gamma-ray spectroscopy; scintillators; energy resolution; nuclear medicine.