SciELO - Scientific Electronic Library Online

 
vol.54 suppl.1Morphological and raman effects on 252Cf irradiated diamondNeutron room return effect índice de autoresíndice de assuntospesquisa de artigos
Home Pagelista alfabética de periódicos  

Serviços Personalizados

Journal

Artigo

Indicadores

Links relacionados

  • Não possue artigos similaresSimilares em SciELO

Compartilhar


Revista mexicana de física

versão impressa ISSN 0035-001X

Resumo

VEGA-CARRILLO, H.R et al. Calculation of Response matrix of a BSS with 6LiI scintillator. Rev. mex. fis. [online]. 2008, vol.54, suppl.1, pp.57-62. ISSN 0035-001X.

The response matrix of a Bonner sphere spectrometer was calculated using MCNP 4C and MCNPX 2.4.0 codes. As thermal neutron detector a 0.4 cm x 0.4 cm 6 LiI which is located at the center of a set of polyethylene spheres. The response was calculated for 0, 2, 3, 5, 8, 10, and 12 inches-diameter polyethylene spheres for neutrons whose energy goes from 2.50E(-8) to 100 MeV. The response matrix was calculated for 23 neutron energies, the response functions were energy-interpolated to 51 neutron energies and were compared with a matrix response reported in the literature, in this comparison both response matrices are in agreement. The main differences were found in the bare detector and are attributed to the irradiation conditions and cross sections, for the other detectors the differences are due to the cross sections libraries.

Palavras-chave : Monte Carlo; 6LiI scintillator; Bonner sphere neutron spectromete.

        · resumo em Espanhol     · texto em Inglês     · Inglês ( pdf )

 

Creative Commons License Todo o conteúdo deste periódico, exceto onde está identificado, está licenciado sob uma Licença Creative Commons