Sistema analógico y digital para el control de un obturador con aplicaciones a la espectroscopia de lente térmica

 

R. Gutiérrez Fuentes1, J.L. Jiménez Pérez1, A. Cruz-Orea2

 

1 Cicata-IPN, Legaria 694, Col. Irrigación, 11500, México D.F.

2 Physics Department, CINVESTAV-IPN, A.P. 14-740, 07360 México D.F., México.

 

Recibido: 31 de octubre de 2006.
Aceptado: 3 de noviembre de 2006.

 

Resumen

Se diseño y construyo un circuito electrónico analógico para controlar un dispositivo electrónico, para el disparo de un obturador (modelo UNIBLITZ VMM-D1). Paralelamente se desarrollo un programa en Labview para el control computarizado de este mismo dispositivo. El circuito analógico proporciona un tiempo de abertura mínimo de 10 ms. Por su parte el programa en Labview nos proporciona un tiempo de abertura de 5 ms y con la opción de regular el número y la frecuencia de los pulsos del obturador. Este obturador es utilizado para proporcionar pulsos de 40 mW de potencia de un haz láser de Ar+, generando gradientes térmicos en muestras líquidas. En este trabajo reportamos la aplicación de este obturador para generar los gradientes térmicos en una solución conteniendo nanopartículas de oro en agua con la finalidad de obtener su difusividad térmica mediante la técnica de lente térmica. Finalmente discutimos las ventajas y desventajas de los dos modos de control.

Palabras clave: Efecto térmico; Técnicas fototérmicas; Líquidos; Instrumentación electrónica.

 

Abstract

It was designed and performed an analogical electronic circuit to control a shutter driver (model UNIBLITZ VMM-D1). Parallel to this it was developed a Labview program for a computerized control of the same device. The analogical circuit provides a minimum opening time of 10 ms. On the other hand the program in Labview provides a minimum time for opening of 5 ms and also with the option to regulate the number and the frequency of the shutter pulses. This shutter is used to provide pulses with 40mW of power from an Ar+ laser beam, generating thermal gradients in liquid samples. In this work we reported the application of this shutter to generate thermal gradients in a solution containing gold nanoparticles in water with the purpose of obtaining its thermal diffusivity by means of the thermal lens technique. Finally we discussed the advantages and disadvantages between the performed shutter controls.

Keywords: Termal effects; Phototermal techniques; Liquids; Electronic Instrumentation.

 

DESCARGAR ARTÍCULO EN FORMATO PDF

 

Referencias

[1] M. Franko, C.D. Tran, Rev. Sci. Instrum. 67, 1 (1996).         [ Links ]

[2] J.C., W. Radolph, Ulltrashort laser pulse phenomena, (Academic Press, San Diego 1996).         [ Links ]

[3] M. A. Garcia Arthur, R. R. Rojo, N. Jamasbi and M. Mohebi, Rev. Méx. Fis. 49, 258 (2003).         [ Links ]

[4] J. P. Gordon, R.C.C. Leite, R.S. Moore, S.P.S. Porto and J.R. Whinnery, J. Appl. Phys. 36, 3 (1965).         [ Links ]

[5] J. Shen, R.D. Lowe and R.D. Snook, Chemical Physics 165, 385 (1992).         [ Links ]

[6] J. F. Sanchez Ramirez, J.L. Jimenez Perez, A. Cruz Orea, R. Gutierrez Fuentes, A. Bautista, and U. Pal, J. of Nanoscience and Nanotechnology, 6, 1 (2006).         [ Links ]

[7] UNIBLITZ, VMM-D1, User Manual, 14-0020, Version 1.4. USA 2002.         [ Links ]

[8] R. Boylestad, Electronica teoria de circuitos. Ed. Prentice Hall 1992.         [ Links ]